Recipes: Plasma Treatment: Difference between revisions

From ProtoFab
Jump to navigation Jump to search
Content deleted Content added
created plasma cleaning recipe table
No edit summary
Line 24: Line 24:
|}
|}


* Running gases in sequence may be done by selecting the option to run Gas A then Gas B. Or you may run a full cycle of Gas A, then a full cycle of Gas B. Running gases in sequence will result in lower vacuum reading.
* You may run gases in a two-sequence plasma cycle (GAS 1/A followed by GAS 2/B) or in individual single gas sequences. Running gases in sequence will result in lower vacuum reading.


* Run multiple cycles if longer plasma time is needed, rather than altering baseline recipe. Recipe settings may only be altered by Protofab Staff
* Run multiple cycles if longer plasma time is needed, rather than altering baseline recipe. Recipe settings may only be altered by Protofab Staff.
* The maximum plasma power is 0.300 kW and will be locked at this setting.
* The maximum plasma power is 0.300 kW and will be locked at this setting.

Revision as of 11:27, 11 September 2026

The following is a baseline recipe for plasma cleaning. Oxygen plasma can be used for cleaning organic contamination, while Argon plasma can clean surface oxide.

Baseline Plasma Cleaning Recipe
Gas Pump down time Purge time Plasma time Vent time Plasma Power
A) Oxygen 80 sec 3 sec 30 sec 35 sec 0.300 kW
B) Argon 80 sec 3 sec 60 sec 35 sec 0.300kW
  • You may run gases in a two-sequence plasma cycle (GAS 1/A followed by GAS 2/B) or in individual single gas sequences. Running gases in sequence will result in lower vacuum reading.
  • Run multiple cycles if longer plasma time is needed, rather than altering baseline recipe. Recipe settings may only be altered by Protofab Staff.
  • The maximum plasma power is 0.300 kW and will be locked at this setting.